Skip to content
Get a Quote + Certificate

Primary service

CMM Calibration and Repair

Request a quotation for performance verification or repair.

Request a quotation

StandardISO/IEC 17025:2017CoverageWestern Pennsylvania and the Upper Ohio Valley

Pricing is set out in writing and sent by email.

Verification, not a slogan

Calibration establishes a result with stated uncertainty against reference standards. Most work ordered as CMM calibration is a performance verification against ISO 10360-2 or ASME B89.4.1, combined with geometric adjustment where the result requires it.

X, Y and Z scales are inspected and cleaned first, with a reader-head check following, before geometry testing and the length measurement error run begin. As-left verification and certificate issue close the visit.

What a visit covers

Each step follows the order a technician actually runs it, from scale inspection through certificate issue.

  • 01

    Scales and reader heads

    X, Y and Z scales are inspected and cleaned; reader heads checked with an oscilloscope and adjusted to spec.

  • 02

    Air bearings and drives

    Air bearing and airline condition is checked before geometry testing proceeds on any axis.

  • 03

    Geometry and length error

    Straightness, linearity and squareness in XY, ZX and ZY, then the ISO 10360-2 length measurement error test at 105 points. A fitted rotary table fourth axis is tested separately to ISO 10360-3.

  • 04

    Probing error

    Form error PFTU tested per ISO 10360-5 against a dedicated test sphere, separate from the probe qualification sphere.

Sequence of a visit

  • 01

    Inspect

    Scales, reader heads, air bearings and airlines are checked before any geometry test runs.

  • 02

    Test

    Geometry, length measurement error and probing error are measured against calibrated artifacts.

  • 03

    Adjust

    Geometric adjustment and a volumetric compensation update follow an out-of-tolerance as-found result, where the controller allows it.

  • 04

    Certify

    As-left results are verified and a certificate is issued stating the standard, uncertainty and decision rule.

Why Pittsburgh CMM Calibration and Repair

  • 01

    As-found recorded first

    No adjustment is made before as-found data is captured and reviewed.

  • 02

    Repair alongside verification

    Probe, scale, air bearing and drive repair is scoped in the same visit where practical.

  • 03

    Relocation re-verification

    Machines moved or crashed are re-verified before being returned to production use.

Coverage across Western Pennsylvania and the Upper Ohio Valley

On-site scheduling covers Western Pennsylvania and the Upper Ohio Valley. Machines outside the standard travel radius may require ship-in arrangements.

Discuss CMM Calibration and Repair Requirements

Describe the machine and the standard the job requires.

Pricing is set out in writing and sent by email.
Get a Quote + CertificateEmail us